Welcome to the website of Artbri Technology Limited !
Laboratory testing equipment consultation hotline
Consulting Hotline (Mr. Liu): +86 173-2848-0575
products
classification

5 ton Sheet & Powder resistance compaction density tester

It can measure the resistivity and thickness of the diaphragm under different pressures with four terminals and two poles.
It can measure the resistivity and compaction density of powder materials under different pressures with four terminals and two poles.
It can measure the resistivity of diaphragm with four probe method
The four-probe method can measure the resistivity and compaction density of powder materials under different pressures

  • Features

Quick and easy switching of various tests
Resistance accuracy can reach micro-ohm level, thickness measurement accuracy can reach sub-micron level
Membrane and powder materials can be tested
Four-probe and 2-pole method are realized on one machine

  • Model and function configuration table

Parameters\Model MFDZ-1TDCR MFDZ-5TDCR MFDZ-1TACR
Pressure kg 1000 5000 1000
Resistance range 0.1uΩ~1000MΩ AC 0.1µΩ~3KΩ
Thickness resolution 0.01um
Thickness accuracy ±0.5um
Resistance accuracy 0.05%FS 0.5%FS
Pressure accuracy 0.5%FS
Sample form Flake, powder, film, block
Four probe Have
Four-wire 2 terminal Have
Sheet resistance Have
Compaction density Have

  • Test Data

1. 1# and 2# powders are the same different batch materials, and the resistivity vs. pressure curve 2 overlaps, but according to the data analysis, the coefficient of variation of 2# is lower, indicating that the process conditions of batch 2# are more stable .

1吨膜片粉末电阻测试机电阻变化曲线 1吨膜片粉末电阻测试机电阻变化数值

2. Diaphragm test data

Resistivity pressure curve at different positions of the diaphragm

膜片不同位置的电阻率压强曲线 膜片不同位置的电阻率压强数值

3. Test data of standard block thickness and resistance repeatability

Repetitive measurement 20 times thickness data evaluation

标准块厚度和电阻重复性测试数据 膜片不同位置的电阻率压强数值

Evaluation of resistance data with 20 repeated measurements

重复测量20次电阻数据评估
  • Some application scenarios

1. It is used for battery formulation design and optimization. By measuring the resistivity of different membranes to evaluate the quality of the formulation and whether the additive content is appropriate, it does not need to be made into a battery for evaluation, which can speed up the design iteration of the battery.
2. Improve the mixing and coating process. The resistivity and thickness of the membrane can be measured to evaluate whether the slurry is evenly dispersed, whether the sedimentation status, and whether the coating and baking are reasonable.
3. Formulate the process time of battery production, such as the relationship between the storage time of the pole piece in the atmosphere and the change of resistivity, so as to determine the allowable storage time of the pole piece in the production line.
4. Measure the resistivity of the solid electrolyte under pressure, and provide directions for optimizing the solid electrolyte.
5. Optimize the production process of powder materials and monitor the production quality of powder materials.